Sign up for our latest offers
Newsletter
By submitting your details you consent to their use in accordance with our Privacy Notice
At Cadmus Products, we prioritise efficient order delivery and offer competitive shipping rates worldwide. Our streamlined process ensures clarity and reliability for all your shipping requirements.
TECHNICAL SPECIFICATIONS
INSTRUMENT:
| Description | Highly modular, multi-gas portable gas monitor for the detection of explosive, oxygen, and toxic gas hazards. Suitable for use in potentially explosive atmospheres, indoors, and outdoors. Fitted with up to 6 sensors, including 3 high-power sensors. Up to 4 x 1 Series and 2 x 4 Series sensors may be selected from over 20 different types for a wide range of safety applications. Available in 3 fundamental configurations of 4- and 6-way cartridge (diffusion), and 6-way cartridge with built-in pump. Can be ordered or configured later as required by users with appropriate training. |
| Alarms and Type | Visual, Vibrating, Audible (95 dB at 30 cm), 3 x Instantaneous, TWA, STEL, Negative Drift, Over Limit, Battery, System Integrity |
| Logged Information | Transmitter (enclosure only): Aluminium alloy: 2.45 kg (5.40 lb). 316 stainless steel: 5.37 kg (11.84 lb) Transmitter with display module: Aluminium alloy: 2.92 kg (6.44 lb). 316 stainless steel: 5.84 kg (12.88 lb) XPIS sensor module with cartridge: 0.80 kg (1.76 lb) XP sensor module with cartridge: 0.69 kg (1.52 lb) |
| System Integrity | Event log: All system and sensor critical events with user acknowledgement where required Data log: System and sensor specific readings at defined interval for 15-120 s (user configurable) Configuration log: Factory default, instrument restore point (user configurable), current settings, with full change history) Calibration history: Full calibration history of a given device up to and including the last 10 calibrations Advanced diagnostics: System, sensor, battery, and user operational-related log for maintenance and lifetime operation assistance |
| Action TAGS | Full boot load, security, and diagnostic check on startup USER tag: Easy identification of users or change of service e.g. “out for repair” ACCESSORY tag: Tag sampling accessories for simple verification of gas type compatibility and set operation automatically BUMP tag: Initiate a BUMP test by scanning a gas cylinder tag with full traceability and 100% validation CALIBRATION tag: Read calibration gas cylinder set up and complete the “guided” calibration routine LOCATION tag: Check an asset name, item number, verify location, or entry to a confined space, all of which can be programmed All action TAGS are AES-128-bit encrypted or higher. User programmed via PC or mobile device software |
| Temperature | Certified: -20°C to 60°C (-4°F to 140°F), operational range is sensor dependent |
| Ingress Protection | Waterproof, independently tested IP66 & IP68 (diffusion), IP65 (pumped) |
| Humidity | Operating conditions are dependent upon sensors. Generally, 0-99% RH (non-condensing) is recommended. |
| Connectivity and Software | USB-C 5V 1-2A device-controlled, Bluetooth® 5. X, NFC 13.56 MHz contactless technology. Supports NTAG424 DNA and DESFire EV2(3) for maximum data security and encryption to avoid cyber-attacks. Operates with Device Configurator Mobile App (iOS, Android), Safety Suite 2.0, and PC Lite (Windows 11). PC Lite is included for fundamental system setup and configuration, and data extraction, at the local point of use. Safety Suite 2.0 is recommended for centralised fleet management and custom report generation across multiple locations. |
| Certifications and Approvals | Global intrinsic safety approvals; suitable for use in Zone 0, 1, 2, Class 1, Division 1 & 2, explosive atmospheres. Approved for indoor, outdoor, onshore, and offshore use by CSA (NRTL) to UL, CSA, IECEX, and ATEX requirements. Approvals pending. Contact Honeywell for the latest availability by geography |
HONEYWELL FLEX 5 SERIES SENSORS
Gas1 Sensing Technology Short Ref. Sensor Format
Flammable (%LEL) Catalytic bead LEL CB F/UF 1 Series
Hydrocarbon (%LEL) Non-dispersive infrared HC LEL NDIR 1 Series
Carbon dioxide (%VOL) Non-dispersive infraread NDIR CO2 1 Series
Volatile Organic
Compounds (ppm)3
Photoionisation (PID) VOC (PID) 4 Series
Oxygen (%VOL) Lead-free, long life O2 1 Series
Ammonia (ppm) Electrochemical NH3 4 Series
Carbon Monoxide (ppm) Electrochemical (low H2 response) CO-H 1 Series
Carbon Monoxide (ppm) Electrochemical CO 1 & 4 Series
Carbon Monoxide/
Hydrogen Sulphide2
Electrochemical (dual toxic) CO+H2S 1 Series
Chlorine (ppm) Electrochemical Cl2 1 & 4 Series
Chlorine Dioxide (ppm) Electrochemical ClO2 4 Series
Ethylene Oxide (ppm) Electrochemical EtO 4 Series
Hydrogen Chloride (ppm) Electrochemical HCl 4 Series
Hydrogen (ppm) Electrochemical H2 4 Series
Hydrogen Cyanide (ppm) Electrochemical HCN 1 & 4 Series
Hydrogen Fluoride (ppm) Electrochemical HF 4 Series
Hydrogen Sulphide (ppm) Electrochemical H2S 1 & 4 Series
Nitric Oxide (ppm) Electrochemical NO 1 & 4 Series
Nitrogen Dioxide (ppm) Electrochemical NO2 1 & 4 Series
Ozone (ppm) Electrochemical O3 4 Series
Phosgene (ppm) Electrochemical COCl2 4 Series
Phosphine (ppm) Electrochemical PH3 4 Series
Sulphur Dioxide (ppm) Electrochemical SO2 1 & 4 Series
By submitting your details you consent to their use in accordance with our Privacy Notice