Sign up for our latest offers
Newsletter
By submitting your details you consent to their use in accordance with our Privacy Notice
At Cadmus Products, we prioritise efficient order delivery and offer competitive shipping rates worldwide. Our streamlined process ensures clarity and reliability for all your shipping requirements.
The Kane 975 features the latest flue gas analyser technology with options for up to 7 gas measurements CO (0-10,000 ppm), CO2, O2, NO, NO2, SO2 & H2S. This product is ideal for larger installations and has built in over-range protection. The intuitive user keypad and buttons with large full colour graphic display make the unit easy to use.
Kane 975 Key Features:
Applications:
Measures:
Calculates:
Fuel Types:
Sensor Options:
Specs | |
---|---|
Flue Temperature | - Range: -50 - 1200°C with suitable probe - Resolution: 0.1°C - Accuracy: ±1°C |
Inlet Temperature | - Range: 0-50°C - Resolution: 0.1°C - Accuracy: ±1.0°C |
Pressure | - Range: 150 mbar - Resolution: 0.1mbar - Accuracy: ±0.5% full scale |
Oxygen | - Range: 0-25% - Resolution: 0.1% - Accuracy: ±0.3% |
Carbon Monoxide H2 compensated | - Range: 0-10000ppm, 10000 - 20000ppm for 15 mins - Resolution: 1ppm - Accuracy: ±5ppm < 100ppm, ±20ppm < 400ppm, ±5% > 400ppm - 2000ppm, ±10% > 2000ppm - 10000ppm |
Carbon Monoxide High Range (optional) |
- Range: 0 - 10% - Resolution: 0.01% - Accuracy: ±5% of reading from 0.1% to 10% |
Hydrogen Sulphide | - Range: 0-200ppm - Resolution: 1ppm - Accuracy: ±5ppm < 100ppm, ±5% > 100ppm |
Nitric Oxide (optional) |
- Range: 0-5000ppm - Resolution: 1ppm - Accuracy: ±5ppm < 100ppm, ±5% > 100ppm |
Nitrogen Dioxide (optional) |
- Range: 0 - 1000ppm - Resolution: 1ppm - Accuracy: ±5ppm < 100ppm, ±10ppm < 500ppm, ±5ppm > 500ppm |
Sulpher Dioxide (optional) |
- Range: 0 - 5000ppm - Resolution: 1ppm - Accuracy: ±5ppm < 100ppm, ±5% > 100ppm |
Carbon Dioxide Infra-red (optional) |
- Range: 0 - 20% - Resolution: 0.1% - Accuracy: ±0.3% reading |
Carbon Dioxide | - Resolution: 0.1% - Accuracy: ±0.3% reading - Range: 0 - 99.9% |
Losses | - Resolution: 0.1% - Accuracy: ±1.0% of reading - Range: 0 - 99.9% |
Efficiency |
- Resolution: 0.1% - Accuracy: ±1.0% of reading - Range: 0 - 120% |
Excess Air | - Resolution: 0.1% - Accuracy: ±0.2% - Range: 0 - 2885.0% |
Temp (Nett) | - Resolution: 1.0°C/F - Accuracy: ±2°C ±0.3% reading - Range: 0 - 1200°C/32 - 2200°F |
CO/CO2 ratio | - Resolution: 0.0001 - Accuracy: ±0.0001 - Range: 0 - 0.9999 |
Poison Index | - Resolution: 0.01% - Accuracy: ±0.01 - Range: 0 - 99.99 |
Pre-programmed Fuels | Natural Gas (x2), Town Gas, Gascor, Light Oil, Heavy Oil, Propane, Butane, Anthracite, Coke, Coal, Kinsale Gas, User Fuel (x5) |
Weight | 1.2kg |
Handset Dimensions | 240mm x 165 x 65 |
Probe Dimensions | Dia 8mm with 285mm long stainless steel shaft, type K thermocouple & 3m long neoprene hose |
Ambient Operating Range | -5°C to +50°C/10% to 90% RH non condensing |
Power Supply (battery charger) | - Input: 110Vac/220 Vac nominal - Output 12 Vac off load |
Battery Life | >6 hours from full charge |
Box Contents |
|
---|
By submitting your details you consent to their use in accordance with our Privacy Notice