Please email / call for availability
Get a quote
Enter your details and we will get back to you as soon as possible.

Product Features
  • Failsafe User-Friendly Interface
  • Environmentally Bulletproof
  • Modular and Serviceable
  • Detcon’s oxygen sensors have a long shelf life and are supported by an industry-leading warranty.
Delivery options

At Cadmus Products we aim to deliver orders efficiently and at highly competitive rates. 

  • We can ship consignments of all shapes and sizes and hazardous content across the UK and worldwide. Most of our orders are sent via DHL, however sometimes we may use other couriers.
  • If you are purchasing a laboratory product online you can find out the price of shipping checkout. If you are obtaining a quote for a product, then we will email you
    with our best shipping prices along with your quote.
  • Please see our delivery page for full information

Teledyne Detcon DM-700-O2 Oxygen Sensor

The Teledyne DM-700-O2 is a non-intrusive “Smart” sensor designed to detect and monitor Oxygen in air over the range of 0-25% by volume using electrochemical sensor technology. The plug-in, field-replaceable sensor cell is temperature-compensated and features oversized gold-plated connections that help prevent corrosion. The Model 700’s rugged framework includes an intrinsically safe electro-polished 316 stainless steel housing with fully encapsulated electronics and dual layer surge protection.

This innovative design virtually eliminates sensor failure due to water ingress, corrosion, vibration, and transient spikes. A primary feature of the Model 700 is embedded intuitive software that simplifies the operator interface by guiding the user through routine calibration, configuration, and fault diagnostic functions using a built-in alpha/numeric display. The Model 700 is equipped with standard analog 4-20mA, and Modbus™ RS-485 outputs. Among its unique features is a wireless option that can be used with Detcon’s SmartWireless® product line. Additional integration options include a Remote Alarm Module (RAM), and HART. Detcon’s oxygen sensors have a long shelf life and are supported by an industry-leading warranty

Failsafe User-Friendly Interface

  • LED Display (With Antiglare Cover)
  • Full Text Display Method
  • Non-intrusive Interface
  • Auto Zero/Auto Span
  • Pre-emptive Fault Diagnostics

Environmentally Bulletproof

  • Electropolished 316SS Construction
  • 100% Epoxy Encapsulated Circuitry
  • Bulletproof I/O Protection
  • Water-Proof, Corrosion-Proof,Vibration-Proof

Modular and Serviceable

  • Modular Design
  • Plug and Play Components
  • Quick Thread Release Sensor Endcap (Allen Wrench only required)
  • Integral Calibration Port
Heading Detail
Sensor Type Continuous diffusion/adsorption, 2-electrode electrochemical cell, Plug-in field replaceable Intelligent Type
Sensor Life 2 1/2 to 3 years typical
Measurement Range 0-25% Oxygen by volume
Accuracy/Repeatability ±1% FS
Response/Clearing Time T90 ≤ 10 seconds
Span Drift <5% of signal per year (in first two years)
Outputs Linear 4-20 mA DC, RS-485 MODBUS-RTU
Electrical Classification Explosion proof, CSA and US (NRTL), Class I, Division 1, Groups B, C, D (Tamb = -40°C to +50°C), Class I, Zone 1, Group IIC, ATEX, II 2 G Ex d ib IIB + H2 T4 Gb (Tamb = -40°C to +50°C)
Ingress Protection NEMA 4X, IP66
Safety Approvals cCSAus, ATEX, CE Marking, SIL2 Certified to IEC 61508
Warranty Plug-in Cell - 2 years; Transmitter - 2 Years
Environmental Specifications Operating Temperature Range: -40°F to +122°F; -40°C to +50°C, Storage Temperature Range: -40°F to +122°F; -40°C to +50°C, Operating Humidity Range: 0-99% RH non-condensing, Operating Pressure Range: Atmospheric ±10%
Mechanical Specifications Dimensions: 7”H x 2.2” Dia.; 178mmH x 65mm Dia. (sensor assembly only), 11”H x 6.1”W x 3.75”D; 280mmH x 155mmW x 96mmD (with junction box), Mounting holes (J-box) 5.5”; 140mm center to center, Weight: 2 lbs; 0.907kg (sensor only), 6 lbs; 2.72kg (w/ aluminum j-box), 9 lbs; 4.08kg (w/stainless steel j-box)
Electrical Specifications Power Input: 11-30 VDC,
Power Consumption: Normal operation = 30mA (<0.75 watt), Maximum = 50mA (1.2 watts),
Inrush Current: 1.67A @ 24V,
RFI/EMI Protection: Complies with EN50270, Analog Output: Linear 4-20mA DC (1,000 ohms max loop load at 24VDC), 0mA All Fault Diagnostics, 2mA In-Calibration, 4-20mA 0-100% full-scale, 22mA Over-range condition,
Serial RS-485 Output: RS-485 Modbus™ RTU,
Baud Rate: 9600 BPS (9600,N,8,1 Half Duplex),
Status Indicators: 4-digit LED display with gas concentration, Full-script menu prompts for AutoSpan, Set-up Options, and Fault Reporting,
Faults Monitored: Loop, Input Voltage, Missing Sensor, Zero, Processor, Memory, Calibration,
Cable Requirements: Power/Analog, 3-wire shielded cable, Maximum distance is 13,300 feet with 14 AWG, Serial Output: 2-wire twisted-pair shielded cable specifically for use with RS-485 installations, Maximum distance is 4,000 feet to last sensor,
I/O Protection: Over-voltage, Miswiring, EMI/RFI Immunity

 

700 Series Model DM 700-O2 Oxygen sensor
   
967-343490-25P_  DM-700-O2 Sensor Assembly (no junction box)
967-343491-25P_  DM-700-O2 Sensor Assembly with Aluminum J-Box 
967-343494-25P_ DM-700-O2 Sensor Assembly with 316 SS J-Box
Write Your Own Review
You're reviewing:Teledyne Detcon DM-700-O2 Oxygen Sensor
Your Rating

Sign up for our latest offers

By submitting your details you consent to their use in accordance with our Privacy Notice